ASTM D5542-94(1999)e1
Historical Standard: ASTM D5542-94(1999)e1 Standard Test Methods for Trace Anions in High Purity Water by Ion Chromatography
SUPERSEDED (see Active link, below)
ASTM D5542
1. Scope
1.1 These test methods cover the determination of trace ([mu]g/L) levels of fluoride, acetate, formate, chloride, phosphate, and sulfate in high purity water using ion chromatography in combination with sample preconcentration. Other anions, such as bromide, nitrite, nitrate, sulfite, and iodide can be determined by this method. However, since they are rarely present in significant concentrations in high purity water, they are not included in this test method. Two test methods are presented and their ranges of application, as determined by a collaborative study, are as follows:
1.2 It is the user's responsibility to ensure the validity of these test methods for waters of untested matrices.
1.3 The common practical range of Test Method A is as follows: chloride, 1 to 100 [mu]g/L, phosphate, 3 to 100 [mu]g/L, and sulfate, 2 to 100 [mu]g/L.
1.4 The common practical range of Test Method B is as follows: fluoride, 1 to 100 [mu]g/L, acetate, 10 to 200 [mu]g/L, and formate, 5 to 200 [mu]g/L.
1.5 The values stated in SI units are to be regarded as the standard.
1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
2. Referenced Documents (purchase separately) The documents listed below are referenced within the subject standard but are not provided as part of the standard.
ASTM Standards
D1066 Practice for Sampling Steam
D1129 Terminology Relating to Water
D1192 Guide for Equipment for Sampling Water and Steam in Closed Conduits
D1193 Specification for Reagent Water
D3370 Practices for Sampling Water from Closed Conduits
D3856 Guide for Management Systems in Laboratories Engaged in Analysis of Water
D4210 Practice for Intralaboratory Quality Control Procedures and a Discussion on Reporting Low-Level Data
D4453 Practice for Handling of High Purity Water Samples
D5810 Guide for Spiking into Aqueous Samples
D5847 Practice for Writing Quality Control Specifications for Standard Test Methods for Water Analysis
Keywords
anion; eluant; ion chromatography; trace; Acetate; Anions content; Bromide (ion) analysis; Chloride analysis--water applications; Eluent/elution; Fluoride (ion) analysis; Formate; High-purity water; Iodide ion (soluble) content; Ion chromatography (IC); Nitrite/nitrate content; Phosphate content--water applications; Sulfate content--water; Sulfite content; Trace elements--water analysis ;
ICS Code
ICS Number Code 71.040.30 (Chemical reagents)
DOI: 10.1520/D5542-94R99E01